The device communicates at 9600 baud, no parity, 8 data bits, and 1 stop bit (9600/8n1).
 
Current settings and measurements are continually streamed over the serial bus. Data is encapsulated in packets of between two and five bytes. The packets are structured as follows:
 
{| border="0" width="30em" class="alternategrey sigroktable"
|-
| style="width: 3em; text-align: center;" | 1
| style="width: 3em; text-align: center;" | 2
| style="width: 3em; text-align: center;" | 3
| style="width: 3em; text-align: center;" | 4
| style="width: 3em; text-align: center;" | 5
|-
| 0xa5
| Token
| colspan="3" | Data
|}
 
The number of data bytes depend on the token. These are the tokens:
 
{| border="0" width="95%" class="alternategrey sigroktable"
|-
! Token
! Description
|-
| 0x0d
| Current measurement, as shown on the display
 
|}


== Resources ==
== Resources ==

Revision as of 00:51, 5 June 2013

CEM DT-8852
CEM DT-8852.png
Status in progress
Connectivity USB
Frequency range 31.5Hz - 8kHz
Measurement range (A) 30dB - 130dB
Resolution 0.1dB
Accuracy (94dB@1kHz) 1.4dB
Frequency weighting A, C
Time weighting F, S
Standards IEC 61672-1 Class 2
Website cem-instruments.com

The CEM DT-8852 is an IEC 61672-1 class 2 compliant sound level meter with USB connectivity.

See CEM DT-8852/Info for more details (such as lsusb -vvv output) about the device.

Hardware

Photos

Protocol

The device communicates at 9600 baud, no parity, 8 data bits, and 1 stop bit (9600/8n1).

Current settings and measurements are continually streamed over the serial bus. Data is encapsulated in packets of between two and five bytes. The packets are structured as follows:

1 2 3 4 5
0xa5 Token Data

The number of data bytes depend on the token. These are the tokens:

Token Description
0x0d Current measurement, as shown on the display